Intended Use - Philips Respironics OptiLife Instrucciones De Uso

Ocultar thumbs Ver también para Respironics OptiLife:
Tabla de contenido

Publicidad

Instructions for Use
OptiLife Nasal Mask

Intended Use

The OptiLife Nasal Mask is intended to provide an interface for application of CPAP or bi-level
therapy to patients. The mask is for single patient use in the home and multi-patient use in
the hospital/institutional environment. The mask is to be used on patients (>66lbs/30kg) for
whom CPAP or bi-level therapy has been prescribed.
Important Note: An exhalation port is built into the OptiLife Nasal Mask so a separate
exhalation port is not required.
Symbols
Warning or
Caution
Does not contain
LATEX
natural rubber latex
Warnings:
This mask is not suitable for providing life support ventilation.
This mask is designed for use with CPAP or bi-level systems recommended by your health care
professional or respiratory therapist. Do not wear this mask unless the CPAP or bi-level system is
turned on and operating properly. Do not block or try to seal the exhalation port. Explanation of
the Warning: CPAP systems are intended to be used with special masks with connectors which
have vent holes to allow continuous flow of air out of the mask. When the CPAP machine is
turned on and functioning properly, new air from the CPAP machine flushes the exhaled air out
through the attached mask exhalation port. However, when the CPAP machine is not operating,
enough fresh air will not be provided through the mask, and exhaled air may be rebreathed.
This warning applies to most models of CPAP systems.
If oxygen is used with the device, the oxygen flow must be turned off when the device is not
operating. Explanation of the Warning: When the device is not in operation, and the oxygen
flow is left on, oxygen delivered into the ventilator tubing may accumulate within the device
enclosure. Oxygen accumulated in the device enclosure will create a risk of fire.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of
an open flame.
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration will vary,
depending on the pressure settings, patient breathing pattern, mask selection, and the leak rate.
This warning applies to most types of CPAP and bi-level machines.
This product may contain chemicals (phthalates) which may cause adverse health effects. If you
are pregnant or breast feeding, contact your healthcare professional before use.
Some users may experience skin redness, irritation, or discomfort. If this happens, discontinue
use and contact your healthcare professional.
At low CPAP or EPAP pressures, the flow through the exhalation port may be inadequate to clear
all exhaled gas from the tubing. Some rebreathing may occur.
Hand wash prior to first use. Inspect the mask for damage or wear (cracking, crazing, tears, etc).
Discard and replace any components as necessary.
If an additional exhalation device is added to the patient circuit, you may need to adjust the
pressure level to compensate for the additional leak of the exhalation device.
Tip
Note
Contains PHT (phthalates)
Consult instructions
for use
ENGLISH
System One Resistance
Control Setting
1