of one (1) minute and until visibly clean. Wipe the Light Shield for a minimum of 30 seconds paying
attention to hard to clean surfaces such as the grommets and grommet/shield interface.
b. Repeat with additional wipes until no visible soil is detected on the wipe. Wipe the surfaces again with
a clean disinfectant wipe and allow to remain wet for the contact time per the disinfectant instructions
for use.
3. Visually inspect the Curing Light Body with magnification lenses, including the Light Shield and LED Light
Attachment, to ensure cleanliness and for any damage or contamination. If any contamination is visible,
repeat the cleaning steps above.
4. If the LED Light Attachment lens remains cloudy, marred, or occluded, repeat steps outlined in #1 above
or replace with a new LED Light Attachment. To verify adequate removal of any soil from the LED Light
Attachment lens, use the Radiometer on the Charging Dock to check for adequate light output.
When cleaning this product, prevent liquids from entering any opening. Wipe carefully so that
!
liquid does not run, and always clean the Curing Light Body with the LED Light Attachment
attached to avoid getting moisture or debris in the Magnetic Socket.
CLEANING THE CHARGING DOCK AND RADIOMETER
Be sure to disconnect the power cord from the Charging Dock before cleaning. The charging
dock should be cleaned before every charge to prevent cross-contamination. Do not spray
!
cleaning solution directly onto components. Instead, spray cleaning solutions on a cloth to
wipe. Do not wipe or spray the electrical contacts.
1. Disconnect the power cord from the Charging Dock.
2. Cleaning of the Charging Dock and Radiometer: Use an EPA-registered, intermediate level quaternary
ammonium compound disinfectant product like CaviCide solution or CaviWipes that contains 20% alcohol
or less.
Note: Spray (or soak) the solution (i.e., CaviCide) onto the wipe until saturated (i.e., wet but not dripping).
Use the saturated wipe to wipe per the instructions below.
a. Wipe the Charging Dock surfaces, including the radiometer, with a clean disinfectant wipe
(e.g, CaviWipes) for a minimum of one (1) minute and until visibly clean.
b. Repeat with additional wipes until no visible soil is detected on the wipe. Wipe the surfaces again with
a clean disinfectant wipe and allow to remain wet for the contact time per the disinfectant instructions
for use.
3. Visually inspect the Charging Dock unit, including the Radiometer Sensor, with magnification lenses
to ensure cleanliness and for any damage or contamination. If any contamination is visible, repeat the
cleaning and disinfection steps above.
Additional Cleaning Cautions:
!
• Do not immerse any components of this product in solutions.
• Use of solutions other than those recommended may damage plastic parts and will void
product warranty.
• Do not use: Denatured alcohol, Lysol™, phenol, bleach, ammonia complex or iodine
complex solutions.*
• Do not use cleaning solutions on the electrical contacts.
*Lysol™ is a registered trademark of Reckitt Benckiser.
PRODUCT SPECIFICATIONS
Type of Information
Light Emitting Diodes:
Output peak wavelength range:
Curing Light Body with LED Light Attachment
dimensions:
Charging Dock dimensions:
Weight:
Power Supply performance:
Specifications
An array of three LEDs die-mounted on a substrate
450 – 470 nanometers
Length: 10.7" (27 cm)
Width: 1.3" (3.3 cm)
Length: 5.8" (14.7 cm)
Width: 3.2" (8.1 cm)
Height: 3.0" (7.6 cm)
Curing Light Body with LED Light Attachment: 7.6oz (215g)
Charging Dock: 16.1 oz (456g)
25 x 10 second cures on a full charge
Charging time: < 70 seconds