Table A- 12 Input/Output Wiring (Tb4) - 7Me39400Al04 Expanded I/O Module - Siemens SITRANS F FUS1010 IP65 NEMA 4X Instrucciones De Servicio

Caudalímetros ultrasónico
Ocultar thumbs Ver también para SITRANS F FUS1010 IP65 NEMA 4X:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Table A- 12
Input/Output Wiring (TB4) - 7ME39400AL04 Expanded I/O Module
Pin#
Signal
1
2
3
4
5
AUX 103+
6
AUX 103-
7
AUX 104+
8
AUX 104-
Note
Auxiliary 4-20mA loops are assigned and spanned under menu control of Vo and PGEN
outputs.
AUX Io3
AUX Io4
Vc: 24 VDC typical (+15 VDC to +30 VDC max) Loop Power
R
: 1000 ohms (max), Loop wire resistance plus user's input load resistance
L
I: 4-20mA
FUS1010 IP65 NEMA 4X & IP66 NEMA 7 Quick Start
Operating Instructions, 01/2013, A5E03086486-AC
No Connection
No Connection
No Connection
No Connection
Isolated Loop Supply
Loop-Powered 4-20mA
Isolated Loop Supply
Loop-Powered 4-20mA
TB4
NC
1
NC
2
NC
3
NC
4
5
V
C
-
I
R
L
6
7
V
C
-
I
R
L
8
Function
Connect +30V max. Loop Supply here
PGEN 1 Data Presented as 4-20mA
Connect +30V max. Loop Supply here
PGEN 2 Data Presented as 4-20mA
Appendix