Wiring Temperature Sensor Board
1. Using a flat-blade screwdriver, loosen Terminal Block TB1 and TB2 screws.
2. Wire the RTD liquid 992EC temperature cable as shown in the table below:
3. For single channel use, wire TB2 as shown in figure above.
4. For dual channel use, connect Channel 2 temperature sensor to TB2.
5. Replace I/O Board and secure with four captive screws paying careful attention to pin
6. Replace Access Cover and finger tighten captive thumbscrew.
FUS1010 IP65 NEMA 4X & IP66 NEMA 7 Quick Start
Operating Instructions, 01/2013, A5E03086486-AC
992EC Series Cable
Wire #1 (Black)
Wire #2 (Orange)
Wire #3 (Brown)
Wire #4 (Red)
Wire #5 GND/SHLD (Blue)
Note
*For cathodically protected pipes, do not attach blue #5 wire at RTD end of cable.
alignment.
Note
TB3 and TB4 are also active analog inputs. See wiring table below.
Pin
TB3
Function
Function
1
AUX. 1 IN
AUX. 3 IN
2
AUX. 1 COM AUX. 3
COM
3
AUX. 2 IN
AUX. 4 IN
4
AUX. 2 COM AUX. 4
COM
TB4
Use
Iin1 Input
Iin1
Common
Iin2 Input
Iin2
Common
3.2 Transmitter Wiring
Terminal TB1
To TB1--1
To TB1--2
To TB1--3
To TB1--4
*To TB1--5
Description
Behavior
Analog
4 to 20mA
current
input
referenced
to meter
ground.
Connecting
Load
Wiring
200Ω
305 meters
(1000 ft.)
Max w/o
factory
approval
27