Sensor Wiring - Siemens SITRANS F FUS1010 IP65 NEMA 4X Instrucciones De Servicio

Caudalímetros ultrasónico
Ocultar thumbs Ver también para SITRANS F FUS1010 IP65 NEMA 4X:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Clamp-on Sensors
Clamp-on style sensors are mounted on the surface of the monitored pipe using series
mounting assemblies. Apply a generous quantity of the thermal couplant provided to the tip
of the sensor and attach it securely to the cleaned pipe surface with the proper mounting
assembly. Temperature measurement anomalies resulting from variations in the ambient
conditions can be minimized by insulating the pipe and sensor after installation.
Figure 3-12
Insert sensors are designed to be used in pipes equipped with Thermowells. These are
spring-loaded, 1/4" diameter sensors with 1/2" NPT integral connection heads, available in
several lengths to accommodate a range of pipe sizes.
Proceed to Commissioning (Page 47).
3.6

Sensor Wiring

Connecting Sensors to the Transmitter
1. Open the transmitter top cover. Using a flat blade screwdriver, remove the Cable Strain
2. Observing the upstream and downstream orientation, attach the UP (upstream) and DN
FUS1010 IP65 NEMA 4X & IP66 NEMA 7 Quick Start
Operating Instructions, 01/2013, A5E03086486-AC
Temperature Sensor Connector Head
Assembly
7ME39600CR 992EC Series Cable
Threaded Pipe Fitting
Thermowell
Insert Temperature Sensor
Relief bracket (see figure below).
(downstream) cables to the sensors and make snug. Attach the other ends to the UP and
DN terminals of the flow meter (see figure below).
Thermal Couplant
Spring Loaded Sensing Element
Pipe Wall
Connecting
3.6 Sensor Wiring
45

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Sitrans f fus1010 ip66 nema 7

Tabla de contenido