Descargar Imprimir esta página

Power Connections; Remove The Cover; High-Voltage Barrier; Remove The High-Voltage Barrier - Hach SC1500 Manual Del Usuario

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 96
• Avoid excessive movement. Transport static-sensitive components in anti-static containers or
packages.
• Wear a wrist strap connected by a wire to earth ground.
• Work in a static-safe area with anti-static floor pads and work bench pads.

3.2.3 Power connections

Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
Electrocution hazard. Always remove power to the instrument before making electrical
connections.
If the controller does not have an installed power cord, connect power with conduit or a power cord.
Refer to the sections that follow to connect power with conduit or a power cord.

3.2.3.1 Remove the cover

Remove the cover as shown in the illustrated steps that follow.

3.2.3.2 High-voltage barrier

High-voltage wiring for the controller is located behind a high-voltage barrier in the controller
enclosure. Do not remove the barrier while power is supplied to the controller. Make sure that the
barrier is installed before power is supplied to the controller.

3.2.3.3 Remove the high-voltage barrier

Remove the high-voltage barrier as shown in the illustrated steps that follow.
12 English
D A N G E R
D A N G E R

Publicidad

Capítulos

loading