The manual, which accompanies this product,
may be revised from time to time and must
therefore be reviewed prior to each procedure
in which the product is used.
The Provox Plug is intended for use with Provox
Voice Prostheses, Provox2 Voice Prostheses
and Provox ActiValve.
Check the device integrity prior to each use
by pulling the strap (fig. 1).

Cleaning and disinfection

Clean the Provox Plug after each use, or if it
is removed during use according to fig. 7-10.
Do not use any water other than drinking
water to clean and rinse the device.
Disinfect the Provox Plug at least once a
day if it is removed (fig. 11) with one of the
following methods:
• Ethanol 70% for 10 minutes
• Isopropylalcohol 70% for 10 minutes
• Hydrogenperoxide 3% for 60 minutes
Caution: Do not clean or disinfect by any
other method as this might cause product
damage and injury.
If the Provox Plug looks dirty or has air dried
in an area with a risk of contamination the
device should be both cleaned and disinfected