Calibration - Grundfos Conex DIA-2Q Manual De Instrucciones

Ocultar thumbs Ver también para Conex DIA-2Q:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 156

11.9 Calibration

1. Press [Cal] to switch to the calibration menu.
– The LED next to [Cal] illuminates.
2. Depending on the set access rights, it may be
necessary to enter the four-digit code number
using the [Up] and [Down] buttons. Confirm using
[OK].
3. Select a parameter for the calibration.
In order to avoid an overdosing, the controllers are
switched off, and the final control elements are
closed.
11.9.1 Peculiarities for the calibration of chlorine
with pH compensation
pH calibration is of prime importance,
since during the calibration of the chlorine
value, the device uses the already
calibrated measured pH value as a
Caution
reference value.
Therefore the pH value must be calibrated
first!
Only then calibrate the chlorine value!
11.9.2 Calibrating the pH value
"Cal"
cal. meas.
pH
value
manual
temp.
cal result
cal. cycle
Fig. 38
Calibrating the pH value
46
GRUNDFOS
temperature
DIN/NIST
of buffer
solution
others
slope/asym. pot.
on
1-100 days
buffer 1/2
automatic reading of
- measuring water
temperature
- measuring signal
and automatic
calibration
cal result:
slope/asym. pot.

Publicidad

Tabla de contenido
loading

Tabla de contenido