Troubleshooting - BRAND Titrette Instrucciones De Manejo

Ocultar thumbs Ver también para Titrette:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Troubleshooting

Problem
Liquid is above the piston
Piston difficult to move
Filling not possible
Filling not possible /
liquid is drawn back into the
titrating tube during filling
Air bubbles in the instrument
Titrating not possible
The volume delivered is
smaller than that indicated
The instrument doesn't
indicate any function
64
Possible cause
Piston leaks
Piston/cylinder assembly is
contaminated or damaged by
crystalline deposits
Filling valve stuck
The discharge valve is
contaminated or the titrating
tube has been damaged
Instrument filled too quickly
Filling tube is loose or dam-
aged
Filling valve is loose or the
seal possibly has not been
inserted
Filling tube does not dip into
the liquid
Recirculation tube is not
mounted or mounted
improperly
Discharge valve stuck
The instrument has not been
completely primed
Seal might not have been
inserted or the filling valve is
loose
Filling valve is blocked or
damaged
Internal error
Corrective action
Perform a cleaning, replace the
piston/cylinder assembly
(see page 56).
Perform a cleaning, replace
the piston/cylinder assembly, if
necessary (see page 56).
Clean the filling valve. If the
valve ball is stuck use a 200 µl
plastic pipette tip to loosen it
(see page 60).
Clean the discharge valve or
exchange the titrating tube
(see page 56).
Fill instrument slowly
Fasten the telescoping filling
tube firmly. If necessary, cut the
tube off approx. 1 cm from the
top or replace it.
Check whether the seal has been
inserted, and fasten the valve
securely with the mounting tool.
Fill up the bottle, or correctly
adjust the length of the telescop-
ing filling tube.
Attach the recirculation tube.
The opening must point outward
toward the bottle wall.
Clean or exchange the titrating
tube with its integrated discharge
valve (see page 56).
Prime the instrument again
(see page 41).
Check whether the seal has been
inserted, and fasten the valve
securely with the mounting tool.
Clean, and if necessary replace
the filling valve (see page 60).
Perform a restart: remove the
batteries, wait 1 minute and then
replace them (see page 61).