Troubleshooting
Problem
Piston is difficult to move
Filling not possible
Instrument does not
dispense liquid
Air bubbles in the instrument
Dispensed volume is too low
Possible cause
- Formation of crystals, dirty
- Volume adjusted to minimum setting
- Sticking filling valve
- Sticking discharge valve
- Reagent with high vapor pressure has been
drawn in too quickly
- Seal not inserted, knurled locking nuts not firmly
connected
- Air not expelled from the instrument
- Filling tube not firmly connected or damaged
- Valves dirty, not firmly installed or damaged
- Backflow tube not connected
- Discharge tube not firmly connected or damaged
- Filling tube not firmly connected or damaged
- Valves not firmly connected or damaged
- 26 -
Corrective action
- Stop dispensing immediately. Loose piston by circular motion,
but do not disassemble. Follow all cleaning instructions
( page 17).
- Set to required volume ( page 12).
- Clean filling valve ( page 19). If necessary replace the valve
with sealing washer.
- Clean discharge valve. If valve ball is stuck use yellow plastic
pipette tip (100 µl) to loosen it. If necessary replace discharge
valve.
- Slowly draw in reagent.
- Double check that seal (K) is in place, and that the knurled
locking nuts at the discharge tube and the SafetyPrime
are firmly seated and finger tight.
- Remove air ( page 11).
- Push the filling tube on firmly. If necessary cut off approx.
1 cm of tube at the upper end and re-connect it or replace
filling tube.
- Cleaning procedure ( page 17). Tighten the valves using
the mounting tool. If necessary, replace the valves and sealing
washers.
- Connect backflow tube ( page 10, N).
- Push the original manufacturer's discharge tube on firmly.
Replace a deformed or damaged discharge tube.
- Cleaning procedure ( page 17. Push the filling tube on firmly.
If this does not suffice, cut off approx. 1 cm of the tube at the
upper end and re-connect it or replace filling tube
( page 31).
- Cleaning procedure ( page 17). Tighten the valves using
the mounting tool. If necessary, replace valves and sealing
washers.
valve
™