Medtronic GUARDIAN LINK Manual De Instrucciones página 14

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 107
Do not twist the tester while it is attached to the transmitter. This will damage the
transmitter.
The tester can be used for one year. If you continue to use the tester for more than one
year, the connector pins inside the transmitter could be damaged, because the tester
cannot continue to provide a waterproof seal. For instructions on how to check the
connector pins, see Inspecting the transmitter connector pins, on page 8.
CAUTION: Only use the green colored tester (MMT-7736L) with the transmitter.
Pockets on the transmitter are visible when connected to the tester. Do
not use any other test plug. Other test plugs are not intended for use
with the transmitter and will damage the transmitter and the tester.
Inspecting the transmitter connector pins
This image is an example of how the connector pins should look.
Look inside the connector opening of the transmitter to make sure that the connector pins
are not damaged or corroded. If the connector pins are damaged or corroded, the
transmitter cannot communicate with the charger or the pump. Contact your local
representative. It may be time to replace your transmitter.
Also look for moisture inside the connector opening. If you see any moisture, allow the
transmitter to dry for at least one hour. Moisture inside the connector opening could
cause the transmitter to not work properly and could cause corrosion and damage over
time.
-8-

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Guardian link 3

Tabla de contenido