Biocompatibility
Applied parts
Operating conditions
Storage conditions
Battery life
Transmitter
frequency
Effective radiated
power (ERP)
Effective isotropic
radiated power (EIRP)
Operating range
Transmitter expected
service life
Transmitter wireless communication
Quality of service
The transmitter and insulin pump connect via smart device connectivity. The
transmitter sends glucose data and system-related alerts to the pump. The pump
verifies the integrity of received data after wireless transmission.
Transmitter: Complies with EN ISO 10993-1
Transmitter
Sensor
Transmitter temperature: 0°C to 45°C (32°F to 113°F)
Caution: When operating the transmitter on a tester in air temperatures
greater than 41°C (106°F), the temperature of the transmitter may exceed
43°C (109°F).
Transmitter relative humidity: 10% to 95% with no condensation
Transmitter pressure: 57.60 kPa to 106.17 kPa (8.4 psi to 15.4 psi)
Charger temperature: 10°C to 40°C (50°F to 104°F)
Charger relative humidity: 30% to 75% with no condensation
Transmitter temperature: -20°C to 55°C (-4°F to 131°F)
Transmitter relative humidity: up to 95% with no condensation
Transmitter pressure: 57.6 kPa to 106 kPa (8.4 psi to 15.4 psi)
Charger temperature: -10°C to 50°C (14°F to 122°F)
Charger relative humidity: 10% to 95% with no condensation
Transmitter: Seven days of continuous glucose monitoring immediately
following a full charge.
Charger: The charger uses one new AAA battery to charge the transmitter.
2.4 GHz band, Bluetooth™* wireless technology (version 4.0)
-12.05 dBm (0.06 mW)
-9.9 dBm (0.1 mW)
Up to 1.8 meters (6 feet) in free-air
The transmitter expected service life is one year depending on patient
usage.