Medtronic GUARDIAN LINK Manual De Instrucciones página 31

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 107
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
Conducted
3 Vrms
disturbances
 150 kHz to 80 MHz
induced by RF fields
 6 Vrms
 ISM bands between
 150 kHz to 80 MHz
Electrical fast
±2 kV
transient/burst
 100 kHz repetition
 IEC 61000-4-4
frequency
Surge
Line to Line: ±0.5 kV,
 IEC 61000-4-5
±1 kV
 Line to Ground: ±0.5
kV, ±1 kV, ±2 kV
Note:
U
is the a.c. mains voltage prior to application of the test level.
T
Voltage dips, short
0% U
interruptions, and
0°, 45°, 90°, 135°, 180°,
voltage variations on
225°, 270°, and 315°)
power supply lines
 0% U
 IEC 61000-4-11
 70% for 25/30 cycles
(at 0°)
 0% for 250/300
cycles
Power frequency
30 A/m
(50/60 Hz) magnetic
field
 IEC 61000-4-8
Proximity fields from
IEC 60601-1-2:2014,
RF wireless
Table 9
communications
equipment
  IEC 61000-4-3
Note:
U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601-1-2:2014
Test Level
; 0.5 cycles (at
T
; 1 cycle (at 0°)
T
-25-
Max foreseeable
use condition per
Environment Guidance
IEC
60601-1-2:2014
Not applicable
Requirement does not
apply to this battery
powered device.
Not applicable
Requirement does not
apply to this battery
powered device.
Not applicable
Requirement does not
apply to this battery
powered device.
Not applicable
Requirement does not
apply to this battery
powered device.
30 A/m
For use in a typical
domestic, commercial, or
hospital environment.
IEC
For use in a typical
60601-1-2:2014,
domestic, commercial, or
Table 9
hospital environment.
Electromagnetic

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Guardian link 3

Tabla de contenido