CAUTION
Certain parts of the EX3 Stepper must be free to move by design. Electrostatic Discharge can occur to these electrically ungrounded parts and disrupt
operation of electronic components inside the EX3 Stepper. The user is cautioned to follow operating precautions within this section.
WARNING
The EX3 Stepper is intended for use by healthcare professionals only. This equipment may cause radio interference or may disrupt the operation of
n
nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the EX3 Stepper or shielding the location.
Immunity Level Justification: This device requires special precautions regarding Electrostatic Discharges (ESD). This device is compliant to +/- 4kV
n
ESD immunity level. Static discharges to the EX3 Stepper can occur and cause the position indicators to disconnect. If the device loses connection,
then an error message will be shown on the display and the encoder position will be disabled. If this occurs the operator must return the EX3 to the
starting position and reinitialize the control program to continue. The procedure can then be restarted.
To reduce the chance of an ESD error occurring, observe the following precautions:
Prior to operating the EX3 Stepper, touch the base of the stepper and procedure table. This should reduce any static buildup and reduce
l
chances of an electrostatic discharge.
Synthetic fabrics and carpets are electrical insulating materials and hold static charge on their surfaces. Use of these materials should be
l
minimized where the EX3 Stepper will be used.
Very low levels of humidity also contribute to static charge buildup. The EX3 Stepper should be used in a temperature/humidity controlled
l
environment where at least 20% relative humidity is maintained.
Use of replacement parts not provided by CIVCO may result in increased emissions or decreased immunity of the EX3 Stepper.
n
The EX3 Stepper should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the EX3 Stepper should be
n
observed to verify normal operation in the configuration in which it will be used.
TABLE 1
GUIDANCE AND MANUFACTURER
The EX3 Stepper is intended for use in the electromagnetic environment specified below. The customer or the user of the EX3 Stepper should assure that it
is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Harmonic
emissions
Not Applicable
IEC 61000-3-2
Voltage
fluctuations/
Not Applicable
flicker emissions
IEC 61000-3-3
'
-
S DECLARATION
ELECTROMAGNETIC EMISSIONS
Electromagnetic environment - guidance
The EX3 Stepper uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
The EX3 Stepper is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.