2. To safely terminate operation, remove USB 2.0 cable from computer.
Validated Treatment Planning Software
MIM Software™ MIM Symphony™
Varian VariSeed™ version 3.0 and higher
Varian Vitesse™ version 8.0.2 and higher
NOTE: Refer to treatment planning software manufacturer for version compatibility.
ADJUSTING THE STEPPER AND STABILIZER
1. Grasp handles of FAM and loosen locking control knob on stabilizer.
®
For Micro-Touch
Stabilizer: Adjust weight by turning weight adjustment knob on stabilizer to achieve desired feel to equipment.
2. Obtain satisfactory transducer position and starting image, tighten locking control knob on stabilizer. After stabilizer is locked into position, locking
control knob should not be used until removal from the patient.
3. Use FAM to optimize position of probe.
4. Install sterile drape and grid, position grid platform appropriately and secure with grid rail locking knobs. Refer to Sterile Drape and Template Grid
instructions for use.
5. To manually set longitudinal scale, loosen drum locking knob and position rotary scale drum at zero. Tighten drum locking knob to secure.
WARNING
Before using assembled equipment, verify needle position is correctly aligned with grid display on the ultrasound monitor. If needle position is not
n
correctly aligned with display, adjust grid platform.
If locking control knob on stabilizer will not hold unit securely in place, please contact CIVCO.
n
Longitudinal and rotational scales are for reference only.
n
ADJUSTING AND SECURING GRID PLATFORM
1. Loosen Vertical Adjustment Screws with the provided hex wrench  (3/32"). Align needle with the grid display on ultrasound monitor by turning the
Vertical Adjustment Knob until the vertical needle position is correctly aligned with display.  Tighten Vertical Adjustment Screws.
2. Loosen Horizontal Adjustment Screw with the provided hex wrench (3/32"). Align needle with the grid display on ultrasound monitor by turning the
Horizontal Adjustment Knob until the horizontal needle position is correctly aligned with display. Tighten Horizontal Adjustment Screw.
WARNING
Stepper is required to be cleaned and disinfected any time grid platform is readjusted and secured to ensure exposed areas of the grid platform are
n
properly decontaminated prior to use. See Reprocessing section for instructions.
III. Computer
I. Vertical Adjustment Screws
II. Vertical Adjustment Knob
III. Horizontal Adjustment Screw
IV. Horizontal Adjustment Knobs
V. Adjustment Slide Block
4
I. EX3 Stepper
II. USB 2.0 Cable
EX3™ Stepper