Grab Sample Options - Hach Polymetron 9611sc PO43– LR Operaciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Option
Description
RESET
Sets the volume of reagent in the reagent bottles to
REAGENT
100% full. IMPORTANT: make sure to push enter,
LEVELS
then select USER PREPARED REAGENTS if the
reagents were prepared in-house. Select HACH
PREPARED REAGENTS if the reagents were
prepared by the manufacturer. This selection is
important for accurate measurements!
RESET
Sets the volume of standard solution in the standard
STANDARD
bottle to 100% full. IMPORTANT: make sure to push
LEVELS
enter, then select USER PREPARED STANDARDS if
the standards were prepared in-house. Select HACH
PREPARED STANDARDS if the standards were
prepared by the manufacturer. This selection is
important for accurate measurements!
PRIME
Starts the flow of all reagents through the tube and valve
REAGENTS
system.

Grab sample options

The grab sample menu allows the user to analyze a specific sample or
standard. Two options are available:
• GRAB SAMPLE IN: This option is used to measure an external
sample or standard.
• GRAB SAMPLE OUT: This option is used to extract a sample directly
from a sample line for external analysis.
Measure a grab sample or standard
Use the grab sample funnel to measure grab samples taken from other
areas in the system or to measure a standard for a calibration
verification.
1. Collect 250–500 mL of the sample or standard in a clean container.
2. Rinse the funnel outside of the unit with the collected sample.
3. Install the funnel again.
4. Select GRAB SAMPLE>GRAB SAMPLE IN.
5. Complete the instructions on the display.
Get a grab sample from the analyzer
Use the grab sample tube to manually dispense a grab sample from one
of the sample sources for external analysis.
The sample source is measured by the analyzer immediately after the
sample is dispensed. The value of the sample source and the job
identification number for the grab sample are shown on the display.
1. Select GRAB SAMPLE>GRAB SAMPLE OUT.
2. Complete the instructions on the display.
3. Put the grab sample out tube in a clean container. The grab sample
tube is located in the lower unit of the cabinet on the left side.
4. Push down on the spigot (upper left corner of the lower
compartment) for the grab sample tube to dispense a grab sample.
Set up the system
The configuration settings can be changed in the SETUP SYSTEM
menu or in the analyzer setup. Refer to
on page 9.
1. Select SETUP SYSTEM.
2. Select one of the options.
Option
Description
MEAS MODE
Changes the mode of the measurement cycle.
Options: interval or continuous (default). The
continuous mode measures approximately every
9 minutes.
EDIT INTERVAL
Changes the interval time when MEAS MODE is set
(conditional)
to interval. Options: 10–240 minutes (default =
15 minutes).
MEAS UNITS
Changes the measurement units that are shown on
the display and in the data log. Options: ppb
(default), ppm, mg/L, µg/L.
Start the analyzer setup
English 11