Table 2 Analyzer data (continued)
Element
Definition
AIR PRESS
Reagent air pressure in the reagent bottles (ideally
3.95 to 4.10 psi)
LED DUTY CYCLE
Depends on condition of colorimeter cell and age of
analyzer (typically 7,200 to 40,000 counts)
HEATER DUTY
Percentage of time the colorimeter heater is on to
CYCLE
maintain a constant 50 °C (122 °F)
SAMPLE FLOW
Approximate sample flow into the colorimeter,
measured during the flush cycle
SAMPLE PRESS 1
Sample pressure before the pre-heater block (ideally
2 to 4.5 psi depending on incoming sample pressure)
SAMPLE PRESS 2
Differential pressure measured across the sample pre-
heater during a colorimeter cell flush. Slightly less than
SAMPLE PRESS1 during the flush cycle and zero
when the flush is off.
REAGENT 1
Level of reagent remaining
REAGENT 2
Level of reagent remaining
REAGENT 3
Level of reagent remaining
STANDARD 1
Level of standard solution remaining
STANDARD 2
Level of standard solution remaining
FAN SPEED
The speed of the ventilation fan.
LEAK COUNTS
Indication of possible fluid leaks (range 0 to 1023). A
count of over 511 indicates a fluid leak
Table 3 Measurement data
Element
Definition
LAST MEAS CHANNEL Last channel measured.
LAST MEAS TIME
Time of last measurement.
LAST ABS
Last absorbance reading.
Table 3 Measurement data (continued)
Element
Definition
LAST CONC
Concentration of last measurement.
NEXT MEAS TIME
Time the next measurement will be taken.
DARK
Number of A/D counts measured when the LED is
turned off.
REF
Reference A/D count used to compensate for natural
color and turbidity.
SAMPLE
A/D counts measurement (after color development)
used to determine the concentration of the sample.
DARK STD DEV
Standard deviation of dark counts out of 6 readings.
REF STD DEV
Standard deviation of reference counts out of
6 readings.
SAMPLE STD DEV
Standard deviation of sample counts out of 6 readings.
SAMPLE VOLUME
Total sample flush volume through the colorimeter for
the measurement cycle.
REAGENT 1
Calculated reagent delivery time to the sample based
on temperature, pressure and viscosity.
REAGENT 2
Calculated reagent delivery time to the sample based
on temperature, pressure and viscosity.
REAGENT 3
Calculated reagent delivery time to the sample based
on temperature, pressure and viscosity.
English 17