Option
Description
SIGNAL AVERAGE
Selects the number of measurements used to
calculate an average measurement (1–5). This
reduces variability in measurements (default = 1) (no
average).
EDIT ANALYZER
Changes the name that is shown on the top of the
NAME
measurement screen (maximum of 16 characters).
EDIT CHANNEL
Changes the name of the sample source that is
NAME
shown on the measurement screen (maximum of
10 characters).
CONFIGURE
Starts or stops measurements for individual sample
SEQUENCER
sources. Sets the measurement order of the sample
(optional)
sources when there is more than one sample
source.
SET DATE & TIME
Sets the analyzer time and date.
DISPLAY SETUP
Changes the language. Adjusts the order
measurements are shown. Adjusts the display
contrast settings.
DISABLE
Stops scheduled maintenance alerts for individual
REMINDERS
components. Options: tubing, stir bar, colorimeter
cell, air filter, stir motor, air relief valve, pinch valve,
air compressor, reagent valves, sample valves,
standard valves, colorimeter LED, fan filter, air
check valve.
MANAGE DEVICES
Installs or removes input modules. Refer to
devices
INSTRUMENT
Shows the analyzer information. Refer to
INFORMATION
instrument information
CALCULATION
Set up variables, parameters, units and formulas for
the analyzer. Refer to
on page 12.
SETUP OUTPUTS
Selects and configures the 4–20 mA setup, the relay
setup and the error hold mode. Refer to
the outputs
12 English
Manage
on page 12 for more information.
View
on page 18.
Set up the calculation
Configure
on page 13 for more information.
Option
Description
SETUP NETWORK
Shows only if a network card is installed. The
(conditional)
network card support is for Modbus, Profibus and
HART.
SECURITY SETUP
Enables or disables the pass code (default =
HACH55).
AIR PURGE
Enables the air purge when an outside air purge is
used on the analyzer (default = off). To use this
feature, make sure that the fan filter plug is installed.
Refer to the documentation that is supplied with the
air purge kit.
RESET DEFAULTS
Sets the configuration to the factory defaults.
Manage devices
Install or remove input modules.
1. Select MANAGE DEVICES.
2. Select an option.
Option
Description
SCAN FOR
The system shows the connected devices.
DEVICES
Note: If no device is connected, the system returns to
the main measurement screen.
DELETE DEVICE
Removes the device when the device is no longer
connected.
Set up the calculation
Set up variables, parameters, units and formulas for the analyzer.
1. Select CALCULATION.
2. Select an option.
Option
Description
SET VARIABLE X
Selects the sensor referring to the variable X.
SET PARAMETER X Selects the parameter referring to the variable X.