Storage And Disposal - Civco AccuCARE Classic Stepper Manual Del Usuario

Tabla de contenido

Publicidad

Classic Stepper
6. Remove Classic Stepper from enzymatic detergent solution and rinse under running tap water for a maximum of 1 minute, but not less than 50
seconds.
7. Dry Classic Stepper with a soft, clean, lint-free cloth.
NOTE:
If there is residual water inside the stepper, compressed air may be used at a maximum of 150 psi to remove remaining water.
n
Prolonged exposure of anodized aluminum and nylon to Enzol
n
WIPING
1. Using a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal wipe,
wipe surfaces to remove gross soil.
Ensure all crevices and grooves have come into contact with wipe.
n
2. Using a new, clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal
wipe, wipe surfaces until visibly wet.
DISINFECTION
1. Unfold a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth
and thoroughly wet surface.
2. Allow treated surface to remain wet for no more than 2 minutes. Use additional wipe or wipes, if necessary, to ensure continuous 2 minutes wet
contact time.
3. Once 2 minute wet contact time has been reached, allow surfaces to air dry.

STORAGE AND DISPOSAL

WARNING
Dispose of single-use components as infectious waste. Reprocess reusable components after each use according to instructions for use.
n
NOTE:
Device should be stored to keep all components together and secure.
n
For questions or to order additional CIVCO products, please call +1 319-248-6757 or 1-800-445-6741 or visit www.CIVCO.com.
n
All product to be returned must be in its original packaging. Contact CIVCO for further instructions as needed.
n
®
Enzymatic Detergent may cause slight discoloration of surfaces.