8. FIXED ANGLE APPROACH:
Using angle selector grip, slide angle indicator window to desired position and push angle selector lock
toward transducer to secure.
NOTE: Ensure angle selector lock is secure in desired position.
FREE ANGLE APPROACH:
Leave angle selector lock in unlocked position and instrument remains in-plane within constraints of guide
angle range.
9. Select appropriate size gauge insert and slide into guide body chute.
BOLT-ACTION QUICK-RELEASE
1. Rotate quick-release tab to open guide insert funnel.
REMOVING GUIDANCE SYSTEM
1. Rotate guide locking lever to unlocked position to remove guide from VerzaLink locating feature of bracket.
2. Remove and dispose of single-use components according to facility infection control policies.
3. Lift bracket latch to remove from transducer. Retain bracket for reprocessing.
REMOVING GENERAL PURPOSE SENSOR FROM BRACKET
1. Pull general purpose sensor away from bracket to remove.
CLEANING, DISINFECTING AND STERILIZING
WARNING
Users of these products have an obligation and responsibility to provide the highest degree of infection
n
control to patients, co-workers and themselves. To avoid cross-contamination, follow infection control
policies established by your facility.
Dispose of single-use components as infectious waste. Clean reusable components after each use.
n
See your ultrasound system's user guide for reprocessing transducer between use.
n
These procedures for reprocessing have been validated for effectiveness and compatibility. Devices could
n
be damaged or cross-contaminated due to improper reprocessing.
Do not gas sterilize or autoclave bracket or general purpose sensor.
n
If bracket or general purpose sensor damage is evident, discontinue use and contact your CIVCO
n
representative.
with Verza™ Guidance System
4
Tracking Bracket