REMOVING GENERAL PURPOSE SENSOR FROM BRACKET
1. Pull general purpose sensor away from bracket to remove.
REPROCESSING
WARNING
Users of this product have an obligation and responsibility to provide the highest degree of infection control
n
to patients, co-workers and themselves. To avoid cross-contamination, follow infection control policies
established by your facility.
Dispose of single-use components as infectious waste. Clean and sterilize reusable components after each
n
use.
See your ultrasound system's user guide for reprocessing transducer between use.
n
Refer to general purpose sensor instruction for reprocessing.
n
This procedure for reprocessing has been validated for effectiveness and compatibility. Device could be
n
damaged or cross-contaminated due to improper reprocessing.
Do not gas sterilize or autoclave bracket or general purpose sensor.
n
If bracket damage is evident, discontinue use and contact your CIVCO representative.
n
Inspect general purpose sensor for damage such as cracks or breakage. If damage is evident, discontinue
n
use of general purpose sensor and contact CIVCO.
CLEANING THE BRACKET
1. After each use, remove bracket from transducer, discarding single-use, disposable components. Remove
visible contaminants from bracket surface using a small, soft instrument brush. Keep bracket from drying out
until complete cleaning can be accomplished.
2. Inspect bracket for damage such as cracks or breakage.
DISINFECTING BRACKET
NOTE: Follow manufacturer's instructions and recommendations for concentration, time of contact and post-process
procedures.
Solution
®
CIDEX
or equivalent 2%
glutaraldehyde solutions
®
CIDEX
OPA or equivalent .55%
ortho-phthalaldehyde solutions
®
Enzol
®
gigasept
FF (new)
DISINFECTANTS AND CLEANING SOLUTIONS TABLE LEGEND
Country of Origin
DE = Germany
US = United States
STORAGE CONDITIONS
Avoid storing product in areas of temperature extremes or in direct sunlight.
n
Store in a cool, dry place.
n
NOTE: Bracket is validated for 5-year product life.
DISPOSAL
WARNING
Dispose of single-use components as infectious waste.
n
NOTE: For questions or to order additional CIVCO products, please call +1 319-248-6757 or 1-800-445-6741
or visit www.CIVCO.com.
with Verza™ Guidance System
Country
of Origin
Manufacturer
®
ASP
Advanced
US
Sterilization Products
®
US
ASP
Advanced
Sterilization Products
®
US
ASP
Advanced
Sterilization Products
Schülke
DE
Disinfectant Type
CL = Cleaner
HLD = High-level disinfectant
6
Tracking Bracket
Qualified
Active
Use
Ingredient
Soak
Glutaraldehyde
Soak
Ortho-
phthalaldehyde
Pre-
Enzymes,
cleaner
surfactant
Soak
Succinic
dialdehyde
Disinfectant
Type
HLD
HLD
CL
HLD