5.9

Cleaning and disinfection by machine

The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
1. Roll up the cable loosely and lay the product carefully in a
suitable rinsing basket. In doing so, make sure the product is
not touching other instruments or parts of instruments.
2. Start a tested program with the following properties:
– Thermal disinfection: 5 to 10 minutes at 90 to 93 °C,
A
≥ 3000.
0
– Final rinse with distilled or deionized water.
– Sufficient product drying.
3. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.

5.10 Check

1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
2. Check the mechanical functionality of the product:
– Can all buttons be pressed?
If damaged, do not use this product!
23