Codman AS 10 Manual Del Usuario página 25

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 128
Swelling or tenderness at
the pump site
Treatment performance
is erratic
Unable to inject bolus
CODFR11091zz01svn.indd 25
CODFR11091zz01svn.indd 25
Pump or catheter
Aspirate fluid at site
system leakage
and verify contents
Infection at site
Aspirate fluid at site
and verify contents
Foreign body rejection
Aspirate fluid at site
phenomena
and verify contents
Twiddler's Syndrome
Disease related
Physician assessment
Intermittent kinking
Use radiography to
of catheters
verify catheter integrity
Extenuating circumstances
Examine the patient history
to determine if the patient
has experienced high or low
altitudes, fevers, etc.
Incorrect needle used
Check the needle to make
sure the bolus needle provided
is being used
The bolus needle is
Check the needle position
positioned incorrectly
The bolus needle is damaged
Inspect the bolus needle
or occluded
Pump is positioned incorrectly
Use radiography to
check pump position and
catheter system
The catheter is bent or occluded
Use radiography to check pump
position and catheter system
Appendix C: Troubleshooting
If aspiration contents show
pump leakage or catheter system
leakage, perform the following:
1. Perform radiography to determine
the source of the leak
2. Perform surgical revision on pump
or catheter system as required
Treat according to
aspiration contents
Treat according to
aspiration contents
Treat accordingly
Treat accordingly
Explain to patient the effects
of high and low altitude, fever,
etc. on pump performance
Use only the bolus needle
provided and retry
Hold needle perpendicular to
the pump and retry
If bolus needle is suspected to
be damaged or occluded, discard
and use a new bolus needle
Treat accordingly
Treat accordingly
Appendix C – 25
9/27/2011 12:26:24 PM
9/27/2011 12:26:24 PM

Publicidad

Tabla de contenido
loading

Tabla de contenido