Hach NA5600 sc Na+ Manual De Instrucciones página 21

Ocultar thumbs Ver también para NA5600 sc Na+:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
• SCHEDULER function (refer to
Option
HOLD OUTPUTS Holds or transfers outputs for the selected channels.
RUN DAYS
START TIME
INTERVAL
DURATION
OFF DELAY
Figure 10 Scheduler function
1 Duration
2 OFF delay
• WARNING function
Option
WARNING LEVEL Sets the relay to on when the selected warning(s) occurs. Push the LEFT
• PROCESS EVENT function
Option
Description
SELECT
Sets the relay to on when the selected process event(s) occurs. Push the LEFT
EVENTS
arrow to select or unselect a checkbox.
MEASURING 1, 2, 3 or 4—Sets the relay to on during the measurement cycle of
Channel 1, 2, 3 or 4.
CALIBRATE—Sets the relay to on during calibration.
SHUTDOWN—Sets the relay to on when in shutdown mode.
STARTUP—Sets the relay to on during the startup cycle.
GRAB SAMPLE—Sets the relay to on during grab sample measurement.
MARK END OF MEASURE—Sets the relay to on for 1 second at the end of each
measurement cycle.
Figure
Description
Sets the days that the relay operates.
Sets the start time.
Sets the time between activation cycles (0 to 999 seconds, default: 0).
Sets the period of time the relay is energized (0 to 999 seconds, default: 0).
Sets the time for additional hold/output time after the relay has been turned
off (0 to 999 seconds, default: 0).
Description
arrow to select or unselect a checkbox.
10)
3 Interval
4 Time (x-axis)
English 21

Publicidad

Tabla de contenido
loading

Tabla de contenido