Site Requirements for the Agilent 2100 Bioanalyzer

Environment

Environment

Your Agilent 2100 Bioanalyzer will work within the specifications of ambient
temperatures and relative humidity described in
Condensation
C AU T I O N
Condensation will damage the electronics.
The Agilent 2100 Bioanalyzer is designed to operate in a controlled electromagnetic
N O T E
environment (EN61326/A1) where RF transmitters such as mobile telephones should not
be used in close proximity.
In order to function properly, do not place the Agilent 2100 Bioanalyzer on a vibrating
N O T E
surface or near vibrating objects.
78
Do not store, ship or use the instrument under conditions where temperature
fluctuations could cause condensation within the instrument.
If your instrument was shipped in cold weather, leave it in its box and allow it to
warm up slowly to room temperature to avoid condensation.
Table 1
on page 80.