When starting the 2100 Expert Software, wait a few seconds until the self-test
procedures for the Agilent 2100 Bioanalyzer are finished. Otherwise the
software may not recognize the Agilent 2100 Bioanalyzer.
Condensation within the instrument
C AU T I O N
Temperature fluctuations could cause condensation within the instrument.
Do not use the Agilent 2100 Bioanalyzer under conditions with temperature
fluctuations.
Let the instrument equilibrate to room temperature for one day to avoid damage of
the electronics.
5

Starting the Agilent 2100 Bioanalyzer

Turning on the Agilent 2100 Bioanalyzer
Connecting the Bioanalyzer via the USB/Serial Cable
Starting the Agilent 2100 Bioanalyzer Software