Guidelines and Manufacturer Declarations / Resistance to Electromagnetic Interference
The type series 7730.65 scale is designated for operations in the electromagnetic environments indicated below.
Customers or users of type 7730.65 scales should ensure that the scale is used in such an environment.
Interference resistance
tests
Electro-static
+ 6 kV contact
Discharges (ESD) law.
discharge (indirect)
IEC 61000-4-2
+ 8 kV
Airborne discharge
Fast transient electrical
+ 2 kV
interference/bursts
for grid cables
law. IEC 61000-4-5
+ 1 kV
for input and output
cables
Surges law. IEC 61000-
+ 1 kV
4-5
Voltage outer
conductor - outer
conductor
+ 1 kV
Voltage outer
conductor - ground
Voltage drops,
< 5% LT
momentary
for ½ period
interruptions and
(> 95 % collapse)
fluctuations of the
power supply law.
40% LT
IEC
for 5 periods
61000-4 -11
(60 % collapse)
70% LT
for 25 periods
(30 % collapse)
< 5% LT
for 5 sec.
(> 95 % collapse)
Magnetic field at the
3 A/m
supply frequency
(50/60 Hz) law. IEC
61000-4-8
NOTE: LT is the mains alternating voltage before the test level is applied
www.soehnle-professional.com
IEC 60601 Test level
Compliance level
+ 6 kV contact
discharge
+ 8 kV
Airborne discharge
+ 2 kV
for grid cables
+ 1 kV
for input and output
cables
+ 1 kV
Voltage outer
conductor - outer
conductor
Not applicable
< 5% LT
for ½ period
(> 95 % collapse)
40% LT
for 5 periods
(60 % collapse)
70% LT
for 25 periods
(30 % collapse)
< 5% LT
for 5 sec.
(> 95 % collapse)
3 A/m
Electromagnetic environment -
Guidelines
Floors should consist of wood or
concrete or be covered with ceramics.
When the floor is covered with synthetic
materials, the relative humidity should be
no less than 30 %.
The quality of the power supply should
correspond to a typical commercial or
hospital environment.
The quality of the power supply should
correspond to a typical commercial or
hospital environment.
The quality of the power supply should
correspond to a typical commercial or
hospital environment.
If the user of the 7730.65 scale demands
continued function even when the
energy supply is interrupted, we
recommend powering the 7730.65 scale
with an uninterruptible power supply or a
battery.
Magnetic fields at mains frequency
should correspond to the typical values
as found in commercial and hospital
environments.