1) Non-Sparking/Energy Limited wiring (Europe) and Dust Ignition
Proof wiring (Europe/International)
Device label (ATEX/IECEx/C-TICK)
The ATEX certificate listed on the device label can be downloaded from the product page of
our website at: www.siemens.com/LR560. Go to Support > Approvals/Certificates.
The IECEx certificate listed on the device label can be viewed on the IECEx website. Go to:
http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the
certificate number IECEx SIR 09.0149X.
• For power demands see
• For wiring requirements follow local regulations.

Instructions Specific to Hazardous Area Installations

• See also
2) Non-incendive and Dust Ignition Proof wiring (US/Canada)
Device label (FM/CSA)
FM/CSA Class 1, Div 2 installation drawing number A5E02795836.
can be downloaded from the product page of our website at: www.siemens.com/LR560 und
Support.
• For power demands see
Instructions Specific to Hazardous Area Installations
(Reference European ATEX Directive 94/9/EC, Annex II, 1.0.6)
The following instructions apply to equipment covered by certificate numbers Sira
09ATEX9356X and Sira 09ATEX4357X:
1)
For use and assembly and details of marking/coding, refer to the main instructions.
2)
The equipment is certified for use as Category 1D, 1/2D and 2D equipment per certificate
Sira 09ATEX9356X and may be used in hazardous zones 20, 21 and 22. The equipment is
also certified for use as Category 3G equipment per certificate Sira 09ATEX4357X and
may be used in hazardous zone 2.
3)
This equipment has a maximum surface temperature of 139 °C (in an 80°C ambient). Refer
to the applicable code of practice for selection of this equipment with respect to specific
dust ignition temperatures.
7ML19985XR81
Temperature De-Rating Curve
Temperature De-Rating Curve
SITRANS LR560 (FF) – QUICK START MANUAL
on page 20.
on page 11.
on page 20.
Page EN-11