Low Calibration Point
Distance from Sensor Reference Point to Low
Calibration Point: usually process empty level.
Range: 0.0000 to 100.000 m
Values
High Calibration Point
Distance from Sensor Reference Point to High
Calibration Point: usually process full level.
Range: 0.0000 to 100.000 m
Values
Wizard Complete
BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wizard menu when Quick
Options
Start is successfully completed.)
To transfer Quick Start values to the device and return to Program menu, press
Then press
to return to Measurement mode.
Device Address
Note:
The address can only be changed from a remote master such as NI-FBUS Configurator
or DeltaV. For further details see Addressing in the manual,
instruments (7ML19985MP01)
To view the unique address of the device on the network:
• In PROGRAM mode, navigate to: Level Meter > Communication (5.) > Device Address (5.2)
to view the device address.
• Press
to return to Measurement mode.
Communications via Foundation Fieldbus
Notes:
• The following instructions assume that the user is familiar with Foundation Fieldbus.
• You will need the full manual to acquire the list of applicable parameters.
• Application Guides for setting up Foundation Fieldbus devices with SIMATIC PDM can be
downloaded from the product page of our website at: www.siemens.com/LR560. Go to
Support > Software Downloads.
AMS Device Manager
AMS Device Manager is a software package used to commission and maintain SITRANS
LR560 and other process devices. Please consult the operating instructions or online help for
details on using AMS Device Manager version 9.0. (You can find more information at
http://www.emersonprocess.com/AMS/.)
Page EN-16
SITRANS LR560 (FF) – QUICK START MANUAL
.
Foundation Fieldbus for Level
7ML19985XR81
(Finish).