Transmitter
PR1710/6x or
PR1710/7x or
PR1711/6x
PTB 02 ATEX 2058
II (2) G [EEx ib] IIC
Weighing platform
IUXS4-......-.....
or
IFXS4-......-.....
KEMA 04ATEX1074 X
II 2G 1D
EEx ib IIC T4
Remarks
1: Parameters as listed in the EC type examination certificate
KEMA 04 ATEX 1074 X
2: All metal parts must be grounded (earthed).
Intrinsically safe
interface
PR1626/60
PTB 02 ATEX 2056
II (2) G [EEx ib] IIC
Weighing platform
IUXS4-......-.....
or
IFXS4-......-.....
KEMA 04ATEX1074 X
II 2G 1D
EEx ib IIC T4
Remarks
1: Parameters as listed in the EC type examination certificate
KEMA 04 ATEX 1074 X
2: All metal parts must be grounded (earthed).
100
Bedienungsanleitung IUXS4
8
Circuit:
Feed circuit
V
16.7 V
o
I
104 mA
o
P
1.47 W
o
Sensor cable connected to the weighing platform (included in std.
equipment supplied): Be sure to exchange for the one listed below!
Sensor cable (recommended: PR6136 or PR136A)
with the following specifications for shielded cable:
L'
1.1 mH/km
C'
250 nF/km
R'
50 ohms/km
To ensure proper operation, cables longer than 500 m should not be used.
For cables longer than 50m, the screen of the cable may not be connected to the
housing of the weighing platform, or the weighing platform and the interface must
be connected with limited resistivity (e.g., by connecting to an equipotential
grounding busbar).
1
1
For gas
For dust
Vi
18 V
18 V
Ii
160 mA
160 mA
Pi
1.5 W
1.5 W
Ci
16nF
---
Li
---
Date
Written by
15.02.06
Klausgrete
Reviewed by
15.02.06
Klausgrete
Released by
15.02.06
Klausgrete
Feed circuit
Circuit:
V
17.4 V
I
159 mA
o
P
1.5 W
o
C
190nF
o
L
o
Sensor cable connected to the weighing platform
(included in std. equipment supplied): maximum length 50 m
Alternative sensor cable:
Maximum length for cable PR6136 or PR6136A: 380 m
500m are permitted if the cable specifications do not exceed the
following: L' = 0.74 mH/km und C' = 375 nF/km
To ensure proper functioning, 500m should not be exceeded.
For cables longer than 50m, the screen of the cable may not be
connected to the housing of the weighing platform, or the weighing
platform and the interface must be connected with limited resistivity (e.g.,
by connecting to an equipotential grounding busbar).
1
1
For gas
For dust
V
18 V
18 V
Ii
160 mA
160 mA
Pi
1.5 W
1.5 W
Ci
16nF
---
Li
---
Date
Written by
15.02.06
Klausgrete
Reviewed by
15.02.06
Klausgrete
Released by
15.02.06
Klausgrete
Gas:
Zones 1, 2
IIC T4; -20°C ... +60°C
Dust: Zones 20, 21, 22
T155°C; -20°C ... +40°C
Connection to PR1710/6x or PR1710/7x or PR1711/6x
Name
Verification of Intrinsic
36116-741-60-A4
Gas:
Zones 1, 2
IIC T4; -20°C ... +60°C
Dust: Zones 20, 21, 22
T155°C; -20°C ... +40°C
Connection to PR1626/60 IIC
Name
Verification of Intrinsic
36116-741-60-A4
Non-
hazardous
area
Hazardous
area
3
Page
5
of
Safety
Rev.
01
Non-
hazardous
area
Hazardous
area
4
Page
5
of
Safety
Rev.
01