Minebea Intec IUXS4 Serie Instalación Y Advertencias De Seguridad página 99

Ocultar thumbs Ver también para IUXS4 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Weighing platform
IUXS4-......-.....
or
IFXS4-......-.....
KEMA 04ATEX1074 X
II 2G 1D
EEx ib IIC T4
1
1
For gas
For dust
Vi
13 V
13 V
Ii
483 mA
425 mA
Pi
2.0 W
2.0W
Ci
16nF
---
Li
---
Remarks
1: Parameters as listed in the EC type examination certificate
KEMA 04 ATEX1074 X
2: Standard version (no temperature sensor connected)
6-wire cable installed flexibly and connected to weighing
platform.
3: All metal parts must be grounded (earthed).
Weighing platform
IUXS4-......-.....
or
IFXS4-......-.....
KEMA 04ATEX1074 X
II 2G 1D
EEx ib IIC T4
1
1
For gas
For dust
Vi
13 V
13 V
Ii
483 mA
425 mA
Pi
2.0 W
2,0W
Ci
16nF
---
Li
---
Remarks
1: Parameters as listed in the EC type examination certificate
KEMA04 ATEX1074 X
2: All metal parts must be grounded (earthed).
Sensor cable connected to the weighing platform
(included in std. equipment supplied): maximum length
50m. Shields ("screens") installed on both ends
Hazardous area
Gas:
Zones 1, 2
IIC T4; -20°C ... +40°C
Dust: Zones 20, 21, 22
T155°C; -20°C ... +40°C
Date
Name
Written by
15.02.06
Klausgrete
Reviewed by
15.02.06
Klausgrete
Released by
15.02.06
Klausgrete
Sensor cable connected to the weighing platform
(included in std. equipment supplied): maximum lengt
50m. Shields ("screens") installed on both ends
Circuit:
Vo
Io
Po
Co
Lo
Hazardous area
Gas:
Zones 1, 2
IIC T4; -20°C ... +40°C
Date
Name
Written by
15.02.06
Klausgrete
Reviewed by
15.02.06
Klausgrete
Released by
15.02.06
Klausgrete
Indicator
CIXS3
KEMA 03ATEX1157 X
2
II 2 G EEx ib IIC T4
Sensor Output
II 1 D IP6X
Vo
12.6 V
Io
350 mA
Po
927 mW
Co
Lo
Lo/Ro
For further connections
(power supplies, data
interfaces, etc.), see
65607-099-60-A4
Connection to CIXS3
Verification of Intrinsic
Safety
36116-741-60-A4
Indicator
h
FCT01-X..
KEMA 00ATEX1012 X
Sensor Output
12.6 V
II 2 G
EEx ib IIC T4
304 mA
1.96 W (FCT01-X)
1.34 W (FCT01-XV1)
400 nF
For further connections
(power supplies, data
interfaces, etc.), see
65607-099-60-A4
Connection to FCT01-X..
Verification of Intrinsic
Safety
36116-741-60-A4
Bedienungsanleitung IUXS4
1
Page
5
of
Rev.
01
2
Page
5
of
Rev.
01
99
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Iuxs4-ce serie

Tabla de contenido