3. Follow the instructions above to set the day, hour and minutes.
4. Once you have set the minutes and pressed enter, the date and
time are set and the time is displayed.
5. If you want to change the date and time, press and hold Date/
Time AN for approx. 3 seconds until the year number starts
flashing. Now you can enter the new values as described above.

Selecting the correct cuff

Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close fitting in the
centre of the upper arm).
Cuff size
for circumference of upper arm
S
17 - 22 cm
M
22 - 32 cm
M - L
22 - 42 cm
L
32 - 42 cm
L - XL
32 - 52 cm
Only use Microlife cuffs.
Contact your local Microlife Service if the enclosed cuff 6 does
not fit.
Connect the cuff to the device by inserting the cuff connector 7
into the cuff socket 3 as far as it will go.

Selecting the user

This device allows to store the results for 2 individual users. In
addition, there is a guest mode in which results are not stored.
Before each measurement, set the user indicator AP for the
intended user: user 1, user 2 or guest mode.
Press and hold the user indicator AP for approx. 2 seconds until
the user symbol starts flashing. Now select user 1, user 2 or
guest mode by pressing the user indicator again. Wait 2-
3 seconds to confirm your selection.
The first person to measure should select user 1.

Selecting standard or AFIB/MAM mode

This device enables you to select either standard (standard single
measurement) or AFIB/MAM mode (automatic triple measure-
ment). To select standard mode, slide the AFIB/MAM switch 8 on
the side of the device downwards to position «1» and to select
AFIB/MAM mode, slide this switch upwards to position «3».

AFIB/MAM mode (highly recommended)

In AFIB/MAM mode, 3 measurements are automatically taken in
succession and the result is then automatically analysed and
4
displayed. Because blood pressure constantly fluctuates, a result
determined in this way is more reliable than one produced by a single
measurement. AF detection is only activated in AFIB/MAM mode.
• When you select the 3 measurements, the MAM-symbol BN
appears in the display.
• The bottom, right hand section of the display shows a 1, 2 or 3 to
indicate which of the 3 measurements is currently being taken.
• There is a break of 15 seconds between the measurements.
A count down indicates the remaining time.
• The individual results are not displayed. Your blood pressure
will only be displayed after all 3 measurements are taken.
• Do not remove the cuff between measurements.
• If one of the individual measurements was questionable, a
fourth one is automatically taken.

4. Taking a Blood Pressure Measurement

Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the
measurement.
2. Sit down on a back-supported chair and relax for 5 minutes.
Keep the feet flat on the floor and do not cross your legs.
3. Always measure on the same arm (normally left). It is recom-
mended that doctors perform double arm measurements on a
patients first visit in order to determine which arm to measure in the
future. The arm with the higher blood pressure should be measured.
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used (marking on the cuff).
• Fit the cuff closely, but not too tight.
• Make sure that the cuff is positioned 2 cm above the elbow.
• The artery mark on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
• Support your arm so it is relaxed.
• Ensure that the cuff is at the same height as your heart.
6. Slide the lock switch AK down to the «unlock» position. Press
the START/STOP button AM to start measuring.
7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
9. During the measurement, the pulse indicator BL flashes in the
display.