Response Format - RADWAG R Balances Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Character
Mass
Unit
Printout format for verified balances:
1
2
3
Stability marker
Character
Mass
Unit

14.2.1. Response Format

On receipt of a command, the balance responds as follows.
first:
XX_A CR LF
XX_I CR LF
XX _ ^ CR LF
XX _ v CR LF
XX _ E CR LF
Where XX stands for command name.
next:
1 – 3
4
Command
Stability marker
Character
Mass
Unit
[space] for positive values
[-] for negative values
9 characters with right justification
3 characters with left justification
4 - 14
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
[space] for positive values
[-] for negative values
11 characters with right justification (contains decimal marks)
3 characters with left justification
command understood and in progress
command understood but not accessible at this moment
command understood but max threshold is exceeded
command understood but min threshold is exceeded
time limit exceeded while waiting for stable measurement result
(time limit is balance characteristic parameter)
5
6
1÷3 characters
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
[space] for positive values
[-] for negative values
9 characters with right justification
3 characters with left justification
15
16 - 18
7
8 – 16
17
- 84 -
19
20
18 - 20
21
22