Printouts Interval For Continuous Transmission; Cooperation With E2R; Printout - RADWAG R Balances Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Continuous transmission in a calibration unit <CAL. UNIT>, independently from currently
chosen weight unit, measurements are transmitted in unit set on a main measuring device
(calibration unit)
Continuous transmission in a current unit <CURRENT UNIT> measurements are transmitted in
a current unit; the unit changes dynamically whenever the user presses Units button in order to
modify the unit on a display.
Continuous transmission off <NONE>
Procedure:
CAUTION!
Continuous transmission may be turned on/off by means of a command sent from a computer (go to
COMMUNICATION PROTOCOL section)

13.1.3. Printouts Interval for Continuous Transmission

Parameter <P4.1.3. INTERVAL> enables setting frequency of printouts for continuous transmission.
Frequency of printouts is set in seconds, wherein the accuracy is 0.1s.
The user may set any time value ranging from 1000 to 0.1 s.
The setting is valid for continuous transmission in calibration unit and in current unit activated by
means of the balance (see parameter P4.2.1.2), it is also valid for continuous transmission activated
by command sent from a computer (see point 15 COMMUNICATION PROTOCOL).

13.1.4. Cooperation with E2R

E2R is a control system, it controls all weighing processes performed on a balance. Upon activation
of this option, operations on some Databases are possible only with the use of a computer (options
are not accessible on a balance). In order to start cooperation with E2R system set <YES> value of
<E2R SYSTEM> parameter.
CAUTION!
<E2R System> parameter activation may be carried out by the device manufacturer or authorized persons,
exclusively. Rememeber to switch continuous transmission on to provide correct coperation with the E2R
system.

13.1.5. Printout

Type of the printout sent to the port chosen for the computer.
Procedure:
− press
button
− enter <P4 DEVICES> menu
− enter menu group <P4.1 COMPUTER>
− enter <P4.1.7 PRINTOUT> parameter
− select printout:
NONE – no printout selected
NSD. PRN. 1÷4 – one of non-standard printouts (see: non-standard printouts
description)
- 78 -