1.10. BALANCE USER TRAINING
The balance should be utilized and supervised only by users who are trained and experienced in
using such type of weighing instruments.
2. TRANSPORT AND STORAGE
2.1. DELIVERY CHECKLIST
Upon delivery it is necessary to check the package and the device, make sure that your package
bears no signs of damage.
2.2. PACKAGE
Keep all package elements should your device be transported in a future. Remember that only
original packaging can be used for shipping purposes. Prior packing uncouple any cables, remove
any separable components (weighing pan, shields, inserts). The device components shall be packed
into an original packaging, thus being protected against potential damage during transportation
3. UNPACKING AND INSTALLATION
3.1. PLACE OF USE AND ASSEMBLING
•
The balance should be stored and used in locations free of vibrations and shakes, free of air
movement and dust,
•
ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C,
•
ambient relative humidity should not exceed 80%,
•
during balance operation, ambient temperature in the weighing room should not change rapidly,
•
the balance should be located on a stable wall console desk or a stable working table which is
not affected by vibrations and distant from heat sources,
•
take special precaution when weighing magnetic objects, as part of the balance is a strong
magnet. Should such loads be weighed, use under-pan weighing option, which removes the
weighed load from area influenced by the balance's magnet. The hook for under-pan weighing
is installed in balance's base,
•
in order to avoid the influence of static electricity on the measurement process, ground the
balance's housing. The grounding bolt is located at the back of the balance's housing.
3.2. UNPACKING
Cut the adhesive tape. Take the device out of the packaging. Open the accessory box, take the
device components out of it.
3.3. STANDARD DELIVERY COMPONENTS LIST
•
balance
bottom insert (AS balance exclusively)
•
centring ring (AS balance exclusively)
•
•
weighing pan, open-work pan for AS with d=0,01/0,1mg exclusively
•
draft shield (AS and PS balances, d=0.001g, exclusively).
•
power supplier
•
user Manual – CD version
- 10 -