2. Electrical And Pneumatic Connections On Side Of Unit - Air Liquide OSIRIS 2 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
14
17
18
24
19
16
13
15
YL024400 - V 2.1 - 10/2012
Changes to settings are only taken into account after the
VALID button
has been pressed.
Screen highlight button "
This button allows highlighting of the screen
visibility of information when using the equipment at night
or in low light levels. Each time the button is pressed the
highlight remains on (for 30 seconds if running on batteries;
until the next time the button is pressed if running off the
mains). If battery charge is low this facility will not be
available.
"
" button
This button validates changed ventilation parameter values.
"
" button, for displaying current patient
parameters (Frequency and I/E ratio)
II. 2. Electrical and pneumatic connections
on side of unit
Oxygen supply
Foolproofed connector according to the national regulations
in force: From 280 kPa (2.8 bar) to 600 kPa (6 bar).
Patient circuit connection
Output cone standardized to ISO 22/15.
Expiratory valve control connection
This is the connection for the tube that pilots inflation and
deflation of the expiratory valve diaphragm. It controls the
PEEP level during the expiratory phase.
External electrical supply connection
Made with a JACK type connector.
When used with a patient who is in recovery room, Air
Liquide Medical Systems strongly recommends using the
cable clamp
on the side of the equipment to ensure
optimum safety against accidental unplugging. The cable
clamp must be fixed on the rear part of the unit approxi-
mately 5 cm from the electrical outlet.
EN
"
for better
5