Xylem SI Analytics TitroLine 5000 Manual De Instrucciones página 116

Ocultar thumbs Ver también para SI Analytics TitroLine 5000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

116
EQ1
The abbreviations used here have the following meaning:
ml:
Total consumption, e.g. for pH Stat
S:
Slope in ml/time (pH Stat)
EQ:
Consumption at the equivalence point 1 and 2 in ml
EP:
Consumption at the end point in ml
B:
Blank value in ml. Mostly determined by way of titration
T:
Titer of the titration solution (e.g. 0.09986)
M:
Mol; mol- or equivalence weight of the sample (e.g. NaCl 58.44)
Factor 1 – 5 conversion factors
F1 - F5
"Weight", weighed-in quantity in g or volume in ml
W
After selecting a formula, please confirm your selection with <ENTER>/<OK>.
The values for the blank value, the titers and factors F1 - F5 can be entered or read from a global memory
(Fig. 67).
Fig. 67
EP1
EP2*T*M*F1/(W*F2)
(EP2-EP1)*T*M*F1/(W*F2)
(F3*EP2-EP1)*T*M*F1/(W*F2)
(F1/W) * EP1 *F2
((F1/W)*(EP2-EP1) * F3-F4)*F5
Calculation of the consumption in the
equivalence or end point.
Formula for the calculation of
concentration of a sample. Direct
titration to 2 EP. Here EP2 (p and m
value)
Formula for the calculation of the
concentration of a sample. Direct
titration to 2 EP. Here calculation of the
difference between EP2-EP1.
Formula for the calculation of the
concentration of a sample. Direct
titration to 2 EP. Here: calculation of the
difference between EP2-EP1, taking
into account a multiplicative factor for
EP2.
Calculation of the des TAC (Total
Anorganic Carbonat reserve)
Calculation of the FOS (Volatile Organic
Acids)
FOS/TAC-value