Replacing the dosing unit

5.2.1
As a rule, the need for replacing the dosing unit occurs only rarely. The dosing unit has to be replaced, if such a
replacement becomes necessary as a result of a defect, or of an inspection of the titration unit.
The dosing unit is equipped with lateral ribs around its circumference, with one of these ribs being in double
design. This double rib serves as a mark for the correct placement of the dosing unit (Fig. 131).
With <ENTER>/<OK> confirm the «Unit size» and select «Dosing unit exchange» (Fig. 124).
The exchange procedure starts directly with any additional warning!
Please take care that the titration tip is placed in a beaker or in the reagent bottle.
Fig. 124
The piston is raised to about 85% (Fig. 125).
Fig. 125
Then you will be prompted to unlock the attachment (Fig. 126).
Fig. 126