Softkey
Sample ID
16
17
Sensor ID
18
DO OptiOx sensor cap lifetime icon
New sensor cap
Sensor cap needs to be replaced
in less than 1 month
19
pH electrode condition criteria (only SG98)
Slope: 95-105%
Offset: ±(0-15)mV
Electrode is in good condition
®
20
ISM
sensor connected
Stability criterion (only SG98)
Strict
22
Warning messages
23
Buffer groups or standards
SevenGo (Duo) pro™/OptiOx™
Sensor cap needs to be replaced
in less than 6 months
Sensor cap needs to be replaced
in less than 2 weeks
Slope: 94-90%
Offset: ±(15-35)mV
Electrode needs cleaning
Medium
Operating the pH/ORP/Ion/DO meter SG98 and DO meter SG9​ ​ 11
Sensor cap needs to be replaced
in less than 3 months
Sensor cap needs to be replaced
in less than 2 days
Slope: 89-85%
Offset: ±(>35)mV
Electrode is faulty
Fast