Calibration Mode; Calibration Reminder; Ph/Ion Measurement Settings (Only Sg98) - Mettler Toledo SevenGo pro Instrucciones De Montaje

Ocultar thumbs Ver también para SevenGo pro:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

When switching from predefined buffer group to customized buffer group, press Save in the table even if no
values have changed.
Ion standards
Concentrations for up to 5 standards with one standard temperature can be defined (see "pH/ion measurement
settings"). Five concentration units are available:
mmol/L
mol/L
ppm
mg/L
%

6.4.2 Calibration mode

Two calibration modes are offered:
Segmented: the calibration curve is made up of linear segments joining the individual calibration points. If
high accuracy is required, the segment method is recommended.
Linear: the calibration curve is determined using linear regression. This method is recommended for
samples with widely varying values.
Note
These settings applies to both pH and Ion calibration.

6.4.3 Calibration reminder

When the calibration reminder is "On", the user is reminded to perform a new calibration after a certain user-
defined interval (maximum 9999 h) has elapsed.
Press READ to save the interval and another screen appears to select calibration expiration date.
Four different time spans can be programmed. In all four cases, a warning message appears that the electrode
should be calibrated.
Immediately
The meter is immediately blocked for measurement when the predefined interval has elapsed.
Reminder + 1h
The meter is blocked for measurement 1 hour after the predefined interval has elapsed.
Reminder + 2h
The meter is blocked for measurement 2 hours after the predefined interval has elapsed.
Continue Reading
The user can continue measuring when the predefined interval has elapsed.

6.5 pH/ion measurement settings (only SG98)

Measurement Resolution
The resolution for pH and mV needs to be set for the display. Up to 3 decimal places can be chosen depending
on the unit of measurement (see table below).
On display
X.XXX
X.XX
X.X
X
In the ion mode, the measurement resolution depends on the concentration and the unit of the measured ion.
Stability Criterion
The Stability icon appears according to the following stability criteria:
22
Menus and settings​ ​
Description
three decimal places
two decimal places
one decimal place
no decimal places
Option
pH
pH
pH, mV
mV
SevenGo (Duo) pro™/OptiOx™

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sevengo duo proOptioxSg9Sg98

Tabla de contenido