Aesculap
TPLO saw GT666
7.6
Manual cleaning with wipe disinfection
Phase
Step
I
Pre-cleaning
II
Cleaning with enzyme solution
III
Intermediate rinse
IV
Drying
V
Wipe disinfection
VI
Final rinse
VII
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*
Validated with enzyme cleaner "Cidezyme Johnson & Johnson"
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
►
Mobilize non-rigid components during cleaning.
►
Clean the product under running water, using a suitable cleaning brush
until all visible residues have been removed from the surfaces.
►
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Phase II
►
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
►
Spray products with a pH neutral enzyme solution, let soak in for at
least 2 minutes and then wipe off.
Phase III
►
Mobilize non-rigid components during cleaning.
►
Rinse product under running tap water for at least 5 minutes.
►
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
►
Contamination should be removed with a lint-free cloth or soft brush
moistened with enzyme cleaner.
►
Flexible components and cannulations should be rinsed for 20 seconds
with the water pistol (cold water, at least 2.5 bar).
10
Acculan 4
®
T
[°C/°F]
RT (cold)
RT (cold)
RT
RT
-
RT (cold)
RT
t
Conc.
Water
[min]
[%]
quality
≥2
-
DW
≥2
0.8
D–W
≥5
-
DW
-
-
-
>1
-
-
0.5
-
FD–W
-
-
-
►
After manual cleaning, check visible surfaces and areas of flexible
components for residues.
►
If necessary, repeat the cleaning process (phase 1 to 3).
Phase IV
►
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Phase V
►
Wipe all surfaces of the product with a single-use disinfecting wipe.
Phase VI
►
Rinse disinfected surfaces after the prescribed reaction time for at
least 1 minute under running demineralized water.
►
Drain any remaining water fully.
Stage VII
►
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Chemicals
Until visually clean
pH-neutral*
–
-
Meliseptol HBV wipes 50 % Propan-1-ol
-
-