5.3 In Case of Equipment Failure

WARNING
If your i600 system is not working properly, or if you suspect that there
is a problem with the equipment:
▪......
... R emove the device from the patient's mouth and discontinue
use immediately.
▪......
Disconnect the device from the PC and check for errors.
▪......
Contact the manufacturer or authorized resellers.
▪......
... M odifications to the i600 system are prohibited by law as they
may compromise the safety of the user, patient, or a third party.

5.4 Hygiene

WARNING
For clean working conditions and patient safety, ALWAYS wear clean
surgical gloves when:
▪......
... H andling and replacing the tip.
▪......
... U sing the i600 scanner on patients.
▪......
... T ouching the i600 system.
The i600's main unit and its optical window should be kept clean at
all times. Before using the i600 scanner on a patient, be sure to:
▪......
... D isinfect the i600 system
▪......
... U se a sterilized tip

5.5 Electrical Safety

16
User Manual
WARNING
▪......
The i600 system is a Class I device.
▪......
... T o prevent electric shock, the i600 system must only be connected
to a power source with a protective earth connection. If you are
unable to insert the i600-supplied plug into the main outlet,
contact a qualified electrician to replace the plug or outlet. Do not
try to circumvent these safety guidelines.
▪......
... T he i600 system only uses RF energy internally. The amount
of RF radiation is low and does not interfere with surrounding
electromagnetic radiation.
▪......
... T here is a risk of electric shock if you attempt to access the inside
of the i600 system. Only qualified service personnel should access
the system.
▪......
... D o not connect the i600 system to a regular power strip or
extension cord as these connections are not as safe as grounded
outlets. Failure to adhere to these safety guidelines may result in
the following hazards
... ...
... : The total short circuit current of all connected equipment may
exceed the limit specified in EN / IEC 60601-1.
... ...
... : The impedance of the ground connection may exceed the limit
specified in EN / IEC 60601-1.
▪......
... D o not place liquids such as beverages near the i600 system and
avoid spilling any liquid on the system.
▪......
... C ondensation due to changes in temperature or humidity can
cause moisture buildup inside the i600 unit, which may damage
the system. Before connecting the i600 system to a power supply,