Clinician Trial Manual
Fitting the Trial Belt to the Patient
1. Determine which side of the patient is best for placement of the Trial Stimulator.
2. Fit the trial belt to the patient with the pocket opening facing the patient's back.
3. Fasten the belt to the patient.
4. If desired, cut off excess length of the trial belt.
5. Place the Trial Stimulator in the belt pocket with the stimulation ON/OFF button facing up.
6. If OR Cable extension(s) were used, ensure that they are no longer connected to the
OR Cable(s).
7. Connect the OR Cable(s) to the Trial Stimulator.
8. Fasten the belt pocket to secure the Trial Stimulator.
1
3
1 – OR Cable
2 – OR Cable Connection to Trial Stimulator
3 – Trial Stimulator
4 – Trial Belt
5 – Lead or Extension
Replacing the Trial Stimulator Battery
1. On the rear of the Trial Stimulator, push in slightly and slide
down the battery compartment cover.
2. Place the two new AA batteries in the slots, matching the
positive (+) and (-) markings in the compartment.
3. Align the battery compartment cover on the case and slide the
cover into position until it snaps closed.