Trial Stimulator Description - Boston Scientific SC-4108 Manual Del Usuario

Ocultar thumbs Ver también para SC-4108:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 13
Clinician Trial Manual
3. While continuing to hold the lead in place, slide the lever to the locked position.
Slide
Slide
4. Repeat this for additional leads (if applicable).
Note: Once the lead is secured in the connector, the stylet can be rotated in, but not removed
from the lead.
5. Connect to the Trial Stimulator. Refer to instructions on page 8.
6. For Intraoperative Testing instructions, proceed to the instructions for Intraoperative Testing
in the appropriate Leads manual.

Trial Stimulator Description

The Trial Stimulator is a temporary external spinal cord stimulator that is connected via the OR
cable(s) to the lead(s) placed along the spinal cord. This external system is used to test how
effective stimulation is in relieving a patient's pain.
Clinician Trial Manual
90655448-03 REV B 7 of 120
Hold
Hold

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sc-4116Sc-5132Sc-6360-32

Tabla de contenido