True RMS Thermal MultiMeter / English
• Resistance and Continuity test
Range
Resolution Accuracy
600.0Ω
0.1Ω
±(0.5%reading + 10dgt)
6.000kΩ
0.001kΩ
60.00kΩ
0.01kΩ
±(0.5%reading + 5digits)
600.0kΩ
0.1kΩ
6.000MΩ 0.001MΩ
60.00MΩ 0.01MΩ
±(2.5%reading + 10dgt)
• Frequency (electronic circuits)
Range
Resolution
Accuracy
40.00Hz~10kHz
0.01Hz~0.001kHz
±(0.5%reading)
• Frequency (electronic circuits)
Range
Resolution
Accuracy
60.00Hz
0.01Hz
600.0Hz
0.1Hz
6.000kHz
0.001kHz
60.00kHz
0.01kHz
±(0.09%rdg+5digits)
600.0kHz
0.1kHz
6.000MHz
0.001MHz
10.00MHz
0.01MHz
Sensitivity: >2Vrms (@20% 80% duty cycle) and f<100kHz;
>5Vrms (@20% 80% duty cycle) and f>100kHz
• Duty Cycle
Range
Resolution
5.0%~95.0%
0.1%
Pulse frequency range: 40Hz÷10kHz, Pulse amplitude:±5V (100us÷100ms)
• Capacity
Range
Resolution
Accuracy
60.00nF
0.01nF
±(1.5%reading+20digits)
600.0nF
0.1nF
±(1.2%reading+8digits)
6.000uF
0.001uF
±(1.5%reading+8digits)
60.00uF
0.01uF
±(1.2%reading+8digits)
600.0uF
0.1uF
±(1.5%reading+8digits)
6000uF
1uF
±(2.5%reading+20digits)
28
Buzzer
Protection against overcharge
>50Ω
1000Vdc/ac rms
Protection against overcharge
1000Vdc/ac rms
Protection against overcharge
1000Vdc/ac rms
Accuracy
±(1.2%reading + 2digits)
Protection against overcharge
1000Vdc/ac rms
19/07/2018 Version No. 001
• Temperature with K-type probe
Range
Resolution
Accuracy (*)
-40.0°C ~ 600.0°C
0.1°C
±(1.5%reading + 3°C)
600°C ~ 1000°C
1°C
-40.0°F ~ 600.0°F
0.1°F
±(1.5%rdg+ 5.4°F)
600°F ~ 1800°F
1°F
(*) Instrument accuracy without probe; Specified accuracy with stable environmental
temperature at ±1°C.
For long-lasting measurements, reading increases by 2°C.
• Reference standards
Safety
IEC/EN61010-1; UL STD. 61010-1
EMC
IEC/EN 61326-1
Insulation
double insulation
Pollution level
2
Overvoltage category
CAT IV 600V, CAT III 1000V
Max operating altitude
2000m (6562ft)
• Power supply
Battery type
1x7.4V rechargeable Li-ION battery, 1500mAh
Battery charger power supply
100/240Vac, 50/60Hz, 12Vdc, 2A
Low battery indication
symbol "
" on the display
Auto Power Off
after 15 - 60min minutes' idling (may be disabled)
Fuses
F10A/1000V, 10 x 38mm (input 10A)
F800mA/1000V, 6 x 32mm (input mA uA)
• Display
Conversion
TRMS
Characteristics
colour TFT, 6000 dots with bargraph
Sampling frequency
3 times/s
8.2.Environment
Environmental conditions for use
Reference temperature
18°C - 28°C (64°F - 82°F)
Operating temperature
5°C - 40°C (41°F - 104°F)
Allowable relative humidity
<80%HR
Storage temperature
-20° - 60°C (-4°F - 140°F)
Storage humidity
<80%HR
19/07/2018 Version No. 001
True RMS Thermal MultiMeter / English
Protection against overcharge
1000Vdc/ac rms