Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The AD226 is intended for use in the electromagnetic environment specified below. The customer or the
user of the
AD226 should assure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
Power frequency (50/60
Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
IEC 60601 Test
level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output
lines
+1 kV differential mode
+2 kV common mode
< 5% UT (>95% dip in
UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in
UT)
for 5 sec
3 A/m
Compliance
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output
lines
+1 kV differential mode
+2 kV common mode
< 5% UT (>95% dip in
UT)
for 0.5 cycle
40% UT (60% dip in UT)
for
5 cycles
70% UT (30% dip in UT)
for
25 cycles
<5% UT
3 A/m
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile.
If
floors are covered with
synthetic material, the
relative humidity should
be greater than 30%.
Mains power quality
should be that of a
typical commercial or
residential environment.
Mains power quality
should be that of a
typical commercial or
residential environment.
(>95% dip in UT)
for 5 sec Mains power
quality should be that of
a typical commercial or
residential environment.
If the user of the AD226
requires continued
operation during power
mains interruptions, it is
recommended that the
AD226 be powered from
an uninterruptable power
supply or its battery.
Power frequency
magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
environment.

Publicidad