Disassembly
1. Grasp the back of the cushion and pull it away from the frame. You will feel a pull as the
magnets separate. i
2. Pinch the upper corner of the headgear fabric and peel it off the top of the wing. h
Slide the fabric completely off of the wing g, and then repeat with the other side of the
headgear.
f
j
7 Health Care Provider & Clinician Information
MRI Safety Information
The Therapy Mask 3100 NC/Therapy Mask 3100 SP is MR Unsafe. Keep it
outside the MRI scan room (Zone IV). It represents a projectile hazard.
Useful Life
The useful life of the mask depends on the use conditions and maintenance (cleaning,
institutional disinfection, and component replacement) of the device. Inspect the mask parts
regularly for damage or wear. Replace components as deemed necessary.
Multi-Patient Use
Clean and disinfect the mask daily and between patients when in use in a clinical setting.
For multi-patient use in the hospital/institutional environment, use the Disinfection Guide
to reprocess the mask between patients. These instructions can be obtained online at
www.philips.com/IFU, or by contacting Philips Respironics Customer Service at 1-800-345-
6443 (USA or Canada), at +1-724-387-4000, or go to www.respironics.com to find your local
customer service contact information.