Setup
The advanced setup mode lets you customize your meter.
4 types parameter are available.
P1.0: temperature parameter setting (t)
P1.1: Change temperature unit (tUt)
P1.2: Normalization temp.(tnr)
P1.3: Temp. Coefficient (tCo)
P2.0: TDS factor setting
P2.1: Setting TDS factor (tds)
P3.0: Reset meter (rSt)
P3.1: Reset
P4.0: Review Calibration Info. (CAL)
P4.1: Review range 1 calibration information
P4.2: Review range 2 calibration information
P1.1: Change temperature unit (tUt):
1. When in measurement mode, press "SET" more
than 2 sec to enter setup mode. Press "q"or "p"
to select P1.0, press "SET" to enter or press "SET"
for more than 2 seconds to quit.
2. Press "p" to select C or F.
3. Press "SET" to confirm or press "SET" more than 2
sec to return to P1.0 without saving.
P1.2: Normalization temp.(tnr):
After saving temp. unit, the meter will automatically
enter normalization temp. adjustment. Press "p" to
change the temperature to 20°C or 25°C. Press "SET"
to confirm and or press "SET" more than 2 sec to
return to P1.0 without saving.
5