P1.3: Temp. Coefficient (tCo)
In temp. coefficient adjustment, press "q"or "p" to
change the temperature coefficient from 0.0 to 4.0.
Press "SET" to confirm or press "SET" more than 2
sec to return to P1.0 without confirming.
P2.1: Setting TDS factor(tds):
In P2.0, press "SET" to enter P2.1. The factor will
flash on LCD. Press "q"or "p" to change factor
from 0.40 to 1.00. Press "SET" to confirm or press
"SET" more than 2 seconds to return to P2. 0
without saving.
P3.1: Reset
After executing, all parameters will be reset to factor
default. The previous calibration information will be
cleared as well.
In P3.0, press "SET" to enter P3.1. Press "p"to select
Y or N. Press "SET" to confirm or press "SET" more
than 2 seconds to return without confirming.
P4.1: Range 1 calibration info:
In P4.0, press "SET" to enter P4.1 and review the
last calibration concentration. If the meter is not
yet calibrated, "---" will display on the LCD. The
calibration information will be override after
recalibration.
P4.2: Range 2 calibration info:
In P4.1, press "p" to enter P4.2 and review the
last calibration concentration. If range 2 is not yet
calibrated, "---" will display on LCD. In P4.1 or P4.2,
press "SET" to return to P4.0.