Control Solution Testing
You should do a control solution test when you are not sure of your test strip
results and want to check that your Reader's built-in meter and test strips are
working properly.
IMPORTANT:
Control solution results should fall within the control solution range printed
on the test strip instructions for use.
Do NOT use control solution past the expiration date. Discard control solution
3 months after opening or on the expiration date printed on the bottle,
whichever comes first. (Example: open April 15, discard July 15; write the
discard date on the side of the bottle.).
The control solution range is a target range for control solution only, not for
your blood glucose results.
The control solution test does not reflect your blood glucose level.
Use only MediSense (low, medium or high) Glucose and Ketone Control
Solution with the Reader's built-in meter.
Check that the LOT number printed on the test strip foil packet and
instructions for use match.
Replace the cap securely on the bottle immediately after use.
Do NOT add water or other liquid to the control solution.
Contact your FreeStyle Libre 2 System provider (pharmacy or mail order
supplier) for how to obtain control solution.
1. From the Home screen, touch the Settings symbol
arrow and touch Control Solution Test.
ART46983-001_rev-A 24-Aug-2022 2:30pm
. Scroll down using the